Recurring concern
Unreliable chest X-ray imaging and interpretation
- Reports that mention this concern
- 5
- Times this concern has been raised
- 11
- Stated actions linked to this concern
- 3
What this concern includes
Includes failures in the dedicated chest X-ray process from acquisition and anatomical marking through labelling, image access, viewing, interpretation, documentation and consideration of the images when these deficiencies can compromise diagnosis or treatment.
What is not included
- Excludes CT, MRI, ultrasound and other imaging modalities unless the assertion explicitly concerns the same chest X-ray process.
- Excludes failures occurring after a chest X-ray has been reliably interpreted, such as downstream referral or treatment failures.
- Excludes generic radiology staffing, IT, communication or documentation deficiencies unless they directly impair chest X-ray acquisition, access, interpretation or clinical use.
- Excludes chest-drain insertion or other procedures unless the asserted deficiency is specifically the associated chest X-ray process.
Trends
Reports over time
Reports mentioning this concern, over time
Reports about this concern issued each year.
* 2026 is projected from reports observed to 7 Sep 2026.
Most frequent recipients
Most frequent recipients
Reports about this concern sent to each recipient.
Reports and published responses 5 reports
See individual reports that have raised this concern, and the actions taken in response
-
Concerns raised1
Delays in official radiologist reporting of chest X-rays showing cardiomegaly
This report raised 5 other concerns. They are not shown here because they do not form part of this recurring concern.
Linked responses
- Mid and South Essex NHS Foundation Trust
- Royal College of Paediatrics and Child Health
Stated actions An action is something a respondent says it has done, is doing, or plans to do in response to the concern raised.2
-
Mid and South Essex NHS Foundation Trust
Complete and formalise the radiology policy review to document how clinical teams can expedite imaging reports for clinical concerns.
-
Mid and South Essex NHS Foundation Trust
Share the updated radiology escalation guidance with paediatric teams and publish it on the Trust intranet.
Respondent positions A position is what a respondent says about the concern when they do not describe a specific action.1
-
Royal College of Paediatrics and Child Health
Formal chest X-ray reporting delays lie outside the respondent’s control.
-
Concerns raised1
Failure to account for the limited sensitivity of upright chest X-rays when excluding perforation
This report raised 7 other concerns. They are not shown here because they do not form part of this recurring concern.
No linked response statementsNo respondent-stated action or position is linked to these concerns in the published data.
-
Concerns raised1
Use of suboptimal chest X-ray imaging after chest injury
This report raised 4 other concerns. They are not shown here because they do not form part of this recurring concern.
No linked response statementsNo respondent-stated action or position is linked to these concerns in the published data.
-
Concerns raised2
Inconsistent use of high-resolution screens for viewing x-rays
Failure to assess the whole x-ray image rather than only expected findings
This report raised 2 other concerns. They are not shown here because they do not form part of this recurring concern.
Linked response
- Stockport NHS Foundation Trust
Stated actions An action is something a respondent says it has done, is doing, or plans to do in response to the concern raised.1
-
Stockport NHS Foundation Trust
Draft guidance identifying high-resolution screens, access arrangements and support across the Trust.
Respondent positions A position is what a respondent says about the concern when they do not describe a specific action.1
-
Stockport NHS Foundation Trust
Standard Emergency Department screens are sufficiently high resolution, and screen resolution was not identified as affecting diagnosis.
-
Concerns raised6
Incorrect interpretation of the cardiac silhouette
Misinterpretation of chest x-rays
Failure to use a lead anatomical marker when taking chest x-rays
Failure by clinicians to identify incorrectly labelled chest x-rays
Failure to consider both chest x-rays
Incorrect labelling of chest x-rays
This report raised 1 other concern. They are not shown here because they do not form part of this recurring concern.
Linked response
- Royal College of Emergency Medicine
Respondent positions A position is what a respondent says about the concern when they do not describe a specific action.1
-
Royal College of Emergency Medicine
Local services should review the environment, clinician training, ultrasound availability and rapid reporting arrangements.
Actions and positions describe what organisations say in published responses. Implementation is not independently verified.