Sandra Danks · Prevention of Future Deaths report

Pin Get email alerts Request correction

Issued 3 Dec 2014•Teesside

Published report and response evidence

This report’s concerns and statements found in recipients’ published responses. Statements are included even where no link to an individual concern is recorded. A link shows an evidence connection; it does not assign responsibility.

View original report
Concerns
1

Raised in this report

Recipients
2

Named on the report

Responses found
1

Of 2 recipients

Stated actions
5

Described in responses

Full report text

This is the full text from the original published report.

Open published report

Concerns and recipient responses

Select any concern, action or position to view the source wording.

Report evidence summary

Concerns raised1

  1. Lack of backup oxygen provision during interruption of the main oxygen apparatus
    Part of recurring concern: Unsafe provision of oxygen therapy

Stated actions An action is something a recipient says it has done, is doing, or plans to do in response to a concern raised.5

  1. Boc Limited

    Continue monitoring current oxygen concentrator procedures.

  2. Boc Limited

    Review policies, procedures and actions concerning oxygen concentrator supply and the incident.

  3. Boc Limited

    Install fixed oxygen tubing to reduce trip and fall risks.

  4. Boc Limited

    Supply backup oxygen cylinders and train home oxygen patients in equipment use and emergency procedures.

  5. Boc Limited

    Service oxygen concentrators and complete associated risk assessments at regular intervals.

Respondent positions A position is what a recipient says about a concern when it does not describe a specific action.3

  1. Boc Limited

    Removed oxygen equipment was found in working order, and BOC was not contacted during the power failure.

    Disputes the concernThe respondent disagreed with part of the concern or the basis for it.
  2. Boc Limited

    Following its comprehensive review, BOC saw no reason at present to take further action, while continuing to monitor procedures.

    No action considered necessaryThe respondent said that no further action was needed.
  3. Boc Limited

    Existing policies, contractual requirements, backup oxygen provision, training, servicing and risk assessments were considered sufficient for patient safety.

    Existing arrangements considered sufficientThe respondent said that existing arrangements were sufficient, so no further action was needed.

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Lack of backup oxygen provision during interruption of the main oxygen apparatus

Wider context from the report

“(1) An interruption in the electricity supply to the main oxygen apparatus stopped the oxygen provision and there was no back up on the main oxygen apparatus to continue to provide oxygen, thus leaving Mrs Danks in a very vulnerable position. ”

Is this part of a recurring concern?

Yes — Unsafe provision of oxygen therapy.

Open source report

Source evidence

How this respondent action was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Continue monitoring current oxygen concentrator procedures.

Verbatim wording from the response

“Following BOC’s comprehensive review of its policies and procedures in respect of the supply of oxygen concentrators and BOC’s actions in respect of the current incident, BOC sees no reason at present to take any further action. BOC shall however continue to monitor its current procedures and make any adjustments and improvements as necessary.”

Source location

2014-0525-Response-by-BOC-Healthcare
Page 3 · response
Published 3 December 2014

Open published response

Source evidence

How this respondent action was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Review policies, procedures and actions concerning oxygen concentrator supply and the incident.

Verbatim wording from the response

“BOC takes patient safety very seriously and has in place robust policies and procedures to ensure the safety of patients using BOC’s Oxygen Concentrators. BOC has conducted a full review of both its general policies and procedures in respect of the supply of oxygen concentrators and also a full review of BOC’s actions in respect of this specific patient since the initial supply of oxygen in July 2013.”

Source location

2014-0525-Response-by-BOC-Healthcare
Page 1 · response
Published 3 December 2014

Open published response

Source evidence

How this respondent action was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Install fixed oxygen tubing to reduce trip and fall risks.

Verbatim wording from the response

“01.08.13 Fixed tubing installation completed to reduce risk of trips and falls”

Source location

2014-0525-Response-by-BOC-Healthcare
Page 2 · response
Published 3 December 2014

Open published response

Source evidence

How this respondent action was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Supply backup oxygen cylinders and train home oxygen patients in equipment use and emergency procedures.

Verbatim wording from the response

“BOC delivers oxygen to home patients as part of its North East NHS contract. BOC has held the contract since 2011 following a public procurement tender process. BOC follows to the letter the requirements of such contract in terms of the specification given by North East NHS for the equipment and services provided and which are determined by the primary caregivers. As part of an oxygen concentrator installation - for long term oxygen therapy - the contract stipulates that BOC will deliver and train the patient on the use of a back-up oxygen cylinder in case of equipment malfunction or power failure. The back-up oxygen cylinder lasts 8 hours at the patient's prescribed flow rate and BOC will attend to the equipment reported fault within that period of time.”

Source location

2014-0525-Response-by-BOC-Healthcare
Page 1 · response
Published 3 December 2014

Open published response

Source evidence

How this respondent action was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Service oxygen concentrators and complete associated risk assessments at regular intervals.

Verbatim wording from the response

“22.10.13 Machine serviced and RA (risk assessment) completed”

Source location

2014-0525-Response-by-BOC-Healthcare
Page 2 · response
Published 3 December 2014

Open published response

Source evidence

How this respondent position was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Removed oxygen equipment was found in working order, and BOC was not contacted during the power failure.

Verbatim wording from the response

“In dealing with this patient, BOC has followed its robust processes and procedures, and is in compliance with the terms of its contract with North East NHS. The oxygen equipment was checked at the required regular intervals and all necessary risk assessments were carried out in line with BOC's contractual obligations. In the particular instance of this patient and the power failure, BOC was not contacted by the patient or a family member at the time of the power failure - had BOC been contacted, the advice given would have been for this patient to use the back-up oxygen cylinder supplied by BOC until power was reconnected. BOC would then have made a visit to replenish the back-up oxygen cylinder.”

Source location

2014-0525-Response-by-BOC-Healthcare
Page 3 · response
Published 3 December 2014

Open published response

Source evidence

How this respondent position was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Following its comprehensive review, BOC saw no reason at present to take further action, while continuing to monitor procedures.

Verbatim wording from the response

“Following BOC’s comprehensive review of its policies and procedures in respect of the supply of oxygen concentrators and BOC’s actions in respect of the current incident, BOC sees no reason at present to take any further action. BOC shall however continue to monitor its current procedures and make any adjustments and improvements as necessary.”

Source location

2014-0525-Response-by-BOC-Healthcare
Page 3 · response
Published 3 December 2014

Open published response

Source evidence

How this respondent position was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Existing policies, contractual requirements, backup oxygen provision, training, servicing and risk assessments were considered sufficient for patient safety.

Verbatim wording from the response

“BOC takes patient safety very seriously and has in place robust policies and procedures to ensure the safety of patients using BOC’s Oxygen Concentrators. BOC has conducted a full review of both its general policies and procedures in respect of the supply of oxygen concentrators and also a full review of BOC’s actions in respect of this specific patient since the initial supply of oxygen in July 2013.”

Source location

2014-0525-Response-by-BOC-Healthcare
Page 1 · response
Published 3 December 2014

Open published response
Back to top

Information checked against the published report and official responses · Data reviewed 7 Sep 2026 · About data quality and limitations

Official responses located
1/2

Data last updated 7 September 2026