PFD report

Dilys Jenkins · Prevention of Future Deaths report

Pin Get email alerts Request correction

Issued 7 Oct 2015•Cardiff and the Vale of Glamorgan

Report record

Published report and response evidence

This page connects the concerns raised in this report with statements found in recipients’ published responses. A link shows a clear evidence connection; it does not assign responsibility.

View original report
Concerns
1

Raised in this report

Recipients
1

Named on the report

Responses found
0

Of 1 recipient

Stated actions
0

Described in responses

Recipients and published responses

Source document

Full report text

This is the full text from the original published report.

Open published report

Concerns and recipient responses

Select any concern, action or position to view the source wording.

Report evidence summary

Concerns raised1

  1. Failure of tracheostomy tube lengths to match increasing patient size
    Part of recurring concern: Unreliable tracheostomy tube management and safety controls
Responses linked to these concerns

Each statement is shown once, even when linked to more than one concern.

No linked response statements

No respondent-stated action or position is clearly linked to these concerns.

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Failure of tracheostomy tube lengths to match increasing patient size

Wider context from the report

“Both ████████ and ████████ were of the view that the tracheostomy manufacturers had not kept pace with the developments in the population towards larger size. The tracheostomy was a size 8 Portex tracheostomy tube which I heard had a length of 75.9 mm ████████ gave evidence that nowadays this length can be inappropriate and that a length of 85.9 mm may be more desirable. They both referred to learned papers which enclose with this Regulation 28 Notice. I have had regard to Chapter 15 in particular of the paper “Major complications of airway management in the UK” (March 2011) and to the article in Anaesthesia 2008, 63, pages 302 – 306 “An investigation into the length of standard tracheostomy tubes in critical care patients”. This recommendation that the length of the tube be increased by 1 cm and the tube redesign to an angle of 110 – 120 degrees to allow optimal tracheal placement. I concluded that the tube will not lie comfortably if its stoma or intra-tracheal length is too short or too long. While the cause of the dislodgment in the case of Dilys Jenkins is unknown the Coroner is concerned that incorrect length may have been a factor. The Coroner is concerned that the tracheostomy industry should be encouraged to change the length of the tracheostomy tube to match the increasing size of the population, and believes the Intensive Care Society is in the best position to influence industry in this regard. ”

Is this part of a recurring concern?

Yes — Unreliable tracheostomy tube management and safety controls.