Recipient

Stryker UK LimitedIncludes reports addressed to a minister or senior office-holder acting for this organisation.

First report 21 Jan 2025•Latest report 21 Jan 2025

Recipient record

Reports, concerns and published responses

Private and voluntary organisations · Private limited company. This page brings together reports naming this recipient and response statements clearly connected to concerns raised in those reports.

Reports
1

Naming this recipient

Published responses
0%

Found for named reports

Concerns addressed
0

Across all linked responses

Stated actions
0

Described in responses

Reports over time

Reports over time

Reports naming this recipient by issue year.

Evidence profile

Report topics

Share of this recipient’s reports compared with all other recipients.

0%published responses found
0stated actions described

Topic comparisons are not available in the current evidence snapshot.

Concerns and recipient responses

Statements from Stryker UK Limited linked to the concerns in each report. Select any concern, action or position to view the source wording.

  1. Addressed to: Managing Director of Stryker (UK) Ltd.

    Dorset

    AI-generated summary

    Reginald Victor Smith · Prevention of Future Deaths report

    This summary was generated using AI from the published report. Please read the original report for the complete account.

    Report summary

    Reginald Victor Smith suffered a witnessed fall at his care home and underwent surgery for a fractured neck of femur. After failure of the metalwork, he underwent revision surgery, received palliative care, and died in hospital on 7 December 2023. Concerns included the possible loosening or deformation of a repeatedly used surgical jig, its loss while sent for manufacturer analysis, and the absence of quality control or auditing of jig integrity before surgery.

    Read the report on judiciary.uk

    Source evidence

    How this individual concern was interpreted

    PFD Monitor created a concise, searchable interpretation from the report wording shown below. The report was sent to Stryker UK Limited; that does not assign responsibility.

    PFD Monitor interpretation

    Lack of auditing and spot checks of surgical jig integrity

    Wider context from the report

    “2) I have concerns with regard to the following: i) Each jig is used many times in surgery having been sterilised after each procedure. It is hammered into the thigh bone and on this occasion may have become deformed over time. ii) The jig was sent away to the manufacturer for analysis but was lost and so no information was available to the court in relation to its integrity. iii) There is no quality control in place in relation to the examination of the jigs being used (other than when it is assembled in theatre by a nurse) prior to surgery. There is no auditing/ spot checks in relation to the integrity of the jigs. ”
    Open source report

    Source evidence

    How this individual concern was interpreted

    PFD Monitor created a concise, searchable interpretation from the report wording shown below. The report was sent to Stryker UK Limited; that does not assign responsibility.

    PFD Monitor interpretation

    Potential deformation of repeatedly used surgical jigs

    Wider context from the report

    “2) I have concerns with regard to the following: i) Each jig is used many times in surgery having been sterilised after each procedure. It is hammered into the thigh bone and on this occasion may have become deformed over time. ii) The jig was sent away to the manufacturer for analysis but was lost and so no information was available to the court in relation to its integrity. iii) There is no quality control in place in relation to the examination of the jigs being used (other than when it is assembled in theatre by a nurse) prior to surgery. There is no auditing/ spot checks in relation to the integrity of the jigs. ”
    Open source report

    Source evidence

    How this individual concern was interpreted

    PFD Monitor created a concise, searchable interpretation from the report wording shown below. The report was sent to Stryker UK Limited; that does not assign responsibility.

    PFD Monitor interpretation

    Lack of preoperative examination of surgical jigs

    Wider context from the report

    “2) I have concerns with regard to the following: i) Each jig is used many times in surgery having been sterilised after each procedure. It is hammered into the thigh bone and on this occasion may have become deformed over time. ii) The jig was sent away to the manufacturer for analysis but was lost and so no information was available to the court in relation to its integrity. iii) There is no quality control in place in relation to the examination of the jigs being used (other than when it is assembled in theatre by a nurse) prior to surgery. There is no auditing/ spot checks in relation to the integrity of the jigs. ”
    Open source report
Back to top

Published response patterns

Compared with other recipients in reports included in PFD Monitor

Describes published response evidence, not performance.

Published responses found

0%
0%All other recipients 58%
0%100%

How actions were described at the time

This respondent
All other recipients
47%25%27%<1%<1%
  • Completed
  • In progress
  • Planned
  • Unclear
  • Partially completed

Statuses reflect what recipients said at the time. PFD Monitor does not verify whether actions happened.

Types of action described in responses

Percentages use all actions described by each group. An action may have more than one type, so percentages do not total 100%.

Information checked against published PFD reports and official responses · Data reviewed 7 Sep 2026 · About data quality and limitations

Data last updated 7 September 2026