PFD report

Najeeb Katende · Prevention of Future Deaths report

Pin Get email alerts Request correction

Issued 21 Apr 2017•Inner North London

Report record

Published report and response evidence

This page connects the concerns raised in this report with statements found in recipients’ published responses. A link shows a clear evidence connection; it does not assign responsibility.

View original report
Concerns
2

Raised in this report

Recipients
1

Named on the report

Responses found
0

Of 1 recipient

Stated actions
0

Described in responses

Recipients and published responses

Source document

Full report text

This is the full text from the original published report.

Open published report

Concerns and recipient responses

Select any concern, action or position to view the source wording.

Report evidence summary

Concerns raised2

  1. Failure to cross-check whether a shockable rhythm is present
  2. Failure to use a defibrillator in AED mode when first attending or when uncertain about heart-rhythm interpretation
    Part of recurring concern: Unreliable defibrillator operation during resuscitation
Responses linked to these concerns

Each statement is shown once, even when linked to more than one concern.

No linked response statements

No respondent-stated action or position is clearly linked to these concerns.

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Failure to cross-check whether a shockable rhythm is present

Wider context from the report

“2. Evidence was also given from other Ambulance staff that: • Despite the presence of other staff between 10.12am and 10.36am, no cross check was made as to whether Najeeb had a shockable rhythm; • If an Automated External Defibrillator, such as those used by members of the public had been applied, this would have detected a shockable rhythm and would have proceeded to defibrilate Najeeb. 3. I consider that it would be of great benefit if LAS were to take the following steps, namely training and instruction to staff: • Actively cross check with another clinician whether a shockable rhythm is present when attending an incident of this sort; • Use the defibrillator in AED mode when first attending as a matter of routine, or at the very least if uncertain when interpreting a heart rhythm; • Further educate on the interpretation of shockable rhythms from readings provided by defibrillator devices. ”

Is this part of a recurring concern?

No recurring-concern membership is currently published.

Open source report

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Failure to use a defibrillator in AED mode when first attending or when uncertain about heart-rhythm interpretation

Wider context from the report

“2. Evidence was also given from other Ambulance staff that: • Despite the presence of other staff between 10.12am and 10.36am, no cross check was made as to whether Najeeb had a shockable rhythm; • If an Automated External Defibrillator, such as those used by members of the public had been applied, this would have detected a shockable rhythm and would have proceeded to defibrilate Najeeb. 3. I consider that it would be of great benefit if LAS were to take the following steps, namely training and instruction to staff: • Actively cross check with another clinician whether a shockable rhythm is present when attending an incident of this sort; • Use the defibrillator in AED mode when first attending as a matter of routine, or at the very least if uncertain when interpreting a heart rhythm; • Further educate on the interpretation of shockable rhythms from readings provided by defibrillator devices. ”

Is this part of a recurring concern?

Yes — Unreliable defibrillator operation during resuscitation.

Open source report
Back to top

Information checked against the published report and official responses · Data reviewed 7 Sep 2026 · About data quality and limitations

Official responses located
0/1

Data last updated 7 September 2026

No official response is included in the current published snapshot.