Recurring concern

Inadequate safety controls for riser-recliner chairs

Pin Get email alerts Request correction

First reported 3 Jul 2017•Latest report 31 Oct 2023

Definition

What this concern includes

Includes failures of controls specifically dedicated to the safe operation of riser-recliner chairs in care settings, including control access restrictions, entrapment detection, user-safety features, risk assessment and associated protective measures where these directly address chair-operation hazards.

Not included

  • Excludes generic care-home risk assessment, staffing, training or equipment-management deficiencies unless they are specifically tied to riser-recliner chair safety.
  • Excludes failures involving other types of chairs, beds, hoists or lifting equipment unless the assertion explicitly concerns riser-recliner chair safety.
  • Excludes general falls, pressure-injury or resident-supervision concerns that do not identify a dedicated riser-recliner chair control failure.
Reports
2

Distinct published reports

Individual concerns
5

A report can raise multiple concerns

Date range
2017–2023

First to latest report issue date

Stated actions
8

Described in published responses

Reports over time

Reports over time

Reports about this concern issued each year.

* 2026 is projected from reports observed to 7 Sep 2026.

Most frequent recipients

Most frequent recipients

Reports about this concern sent to each recipient.

A. J. Way and Co. Ltd1
Care Quality Commission1
Department of Health and Social Care1
Health and Safety Executive1
National Trading Standards1
Recipient name withheld1
Yorkshire Care Equipment1

Concerns and responses across reports

Only concerns grouped under this recurring concern are included. Select any concern, action or position to view the source wording.

  1. Somerset

    AI-generated summary

    Evelyn Ann Burcham · Prevention of Future Deaths report

    This summary was generated using AI from the published report. Please read the original report for the complete account.

    Report summary

    Evelyn Ann Burcham, who had dementia and required full care, was tipped from a riser-recliner chair after another resident accessed and operated its remote control while it was unattended. She sustained a severe brain bleed and died ten days later. The principal concern was that the risk of cognitively impaired residents accessing and operating chair controls had not been foreseen, and that the chairs lacked a suitable safety feature to restrict access to the controls.

    Read the report on judiciary.uk

    Source evidence

    How this individual concern was interpreted

    PFD Monitor created a concise, searchable interpretation from the report wording shown below.

    PFD Monitor interpretation

    Unavailability of riser-recliner chairs with controls that limit unauthorised use

    Wider context from the report

    “I am concerned that these chairs are common place in care home settings where residents suffer from cognitive impairment. The risks of another resident with dementia accessing the controls of the riser-recliner chair(s) was not foreseen by the Home and so was not factored into any operative risk assessment at the time; hence no measures were in place to minimise the risk. There remains a real and immediate risk that those with a cognitive impairment that manifests itself in a compulsion to fiddle and/or press buttons, can create a risk of death to other residents in the same care facility or setting. If this particular care home group did not foresee the risk then it is likely that others have not foreseen it either. I was told at the Home had made enquiries with manufacturers of these standard riser-recliner chairs about the ability to ‘lock’ the remote control and/or find a safe was of storing it so that it is not accessible to anyone who does not have the authority, training or appropriate cognitive function to be able to use it safely. It would appear that a chair with a ‘safe’ remote cannot be purchased by the Home and there do not appear to be any regulatory or manufacturing standards (over and above manufacturing standards for consumers) that require these types of chairs to have this, or some alternative, safety feature that limits the use of the controls. I was told that the only way to render the remote ‘safe’ was to turn the power off at the mains, which in itself could create health and safety issues if the chair needed to be operated quickly. ”

    Source location

    Evelyn Ann Burcham · Prevention of Future Deaths report
    Page 3 · concerns

    Open source report

    Source evidence

    How this individual concern was interpreted

    PFD Monitor created a concise, searchable interpretation from the report wording shown below.

    PFD Monitor interpretation

    Failure to assess and mitigate the risk of cognitively impaired residents accessing riser-recliner chair controls

    Wider context from the report

    “I am concerned that these chairs are common place in care home settings where residents suffer from cognitive impairment. The risks of another resident with dementia accessing the controls of the riser-recliner chair(s) was not foreseen by the Home and so was not factored into any operative risk assessment at the time; hence no measures were in place to minimise the risk. There remains a real and immediate risk that those with a cognitive impairment that manifests itself in a compulsion to fiddle and/or press buttons, can create a risk of death to other residents in the same care facility or setting. If this particular care home group did not foresee the risk then it is likely that others have not foreseen it either. I was told at the Home had made enquiries with manufacturers of these standard riser-recliner chairs about the ability to ‘lock’ the remote control and/or find a safe was of storing it so that it is not accessible to anyone who does not have the authority, training or appropriate cognitive function to be able to use it safely. It would appear that a chair with a ‘safe’ remote cannot be purchased by the Home and there do not appear to be any regulatory or manufacturing standards (over and above manufacturing standards for consumers) that require these types of chairs to have this, or some alternative, safety feature that limits the use of the controls. I was told that the only way to render the remote ‘safe’ was to turn the power off at the mains, which in itself could create health and safety issues if the chair needed to be operated quickly. ”

    Source location

    Evelyn Ann Burcham · Prevention of Future Deaths report
    Page 2 · concerns

    Open source report

    Source evidence

    How this individual concern was interpreted

    PFD Monitor created a concise, searchable interpretation from the report wording shown below.

    PFD Monitor interpretation

    Health and safety risk from disabling riser-recliner chair controls at the mains

    Wider context from the report

    “I am concerned that these chairs are common place in care home settings where residents suffer from cognitive impairment. The risks of another resident with dementia accessing the controls of the riser-recliner chair(s) was not foreseen by the Home and so was not factored into any operative risk assessment at the time; hence no measures were in place to minimise the risk. There remains a real and immediate risk that those with a cognitive impairment that manifests itself in a compulsion to fiddle and/or press buttons, can create a risk of death to other residents in the same care facility or setting. If this particular care home group did not foresee the risk then it is likely that others have not foreseen it either. I was told at the Home had made enquiries with manufacturers of these standard riser-recliner chairs about the ability to ‘lock’ the remote control and/or find a safe was of storing it so that it is not accessible to anyone who does not have the authority, training or appropriate cognitive function to be able to use it safely. It would appear that a chair with a ‘safe’ remote cannot be purchased by the Home and there do not appear to be any regulatory or manufacturing standards (over and above manufacturing standards for consumers) that require these types of chairs to have this, or some alternative, safety feature that limits the use of the controls. I was told that the only way to render the remote ‘safe’ was to turn the power off at the mains, which in itself could create health and safety issues if the chair needed to be operated quickly. ”

    Source location

    Evelyn Ann Burcham · Prevention of Future Deaths report
    Page 3 · concerns

    Open source report

    Source evidence

    How this individual concern was interpreted

    PFD Monitor created a concise, searchable interpretation from the report wording shown below.

    PFD Monitor interpretation

    Risk of death to other residents from cognitively impaired residents operating riser-recliner chair controls

    Wider context from the report

    “I am concerned that these chairs are common place in care home settings where residents suffer from cognitive impairment. The risks of another resident with dementia accessing the controls of the riser-recliner chair(s) was not foreseen by the Home and so was not factored into any operative risk assessment at the time; hence no measures were in place to minimise the risk. There remains a real and immediate risk that those with a cognitive impairment that manifests itself in a compulsion to fiddle and/or press buttons, can create a risk of death to other residents in the same care facility or setting. If this particular care home group did not foresee the risk then it is likely that others have not foreseen it either. I was told at the Home had made enquiries with manufacturers of these standard riser-recliner chairs about the ability to ‘lock’ the remote control and/or find a safe was of storing it so that it is not accessible to anyone who does not have the authority, training or appropriate cognitive function to be able to use it safely. It would appear that a chair with a ‘safe’ remote cannot be purchased by the Home and there do not appear to be any regulatory or manufacturing standards (over and above manufacturing standards for consumers) that require these types of chairs to have this, or some alternative, safety feature that limits the use of the controls. I was told that the only way to render the remote ‘safe’ was to turn the power off at the mains, which in itself could create health and safety issues if the chair needed to be operated quickly. ”

    Source location

    Evelyn Ann Burcham · Prevention of Future Deaths report
    Page 2 · concerns

    Open source report

    Source evidence

    How this respondent action was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Process all future riser/recliner chair orders through Shackleton’s and fit supplied chairs with lockable handsets.

    Verbatim wording from the response

    “Following receipt of the request, Aria Care reached out to Shackleton’s, our furniture supply partner seeking advice and guidance on riser/recliner chairs and the potential for purchase and supply of handsets that could be locked in a secure position and avoid unauthorised access and repositioning.”

    Source location

    Response from Aria Care
    Page 1 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent action was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Work with residents and families to replace privately owned riser/recliner chairs lacking lockable handsets.

    Verbatim wording from the response

    “Where residents are currently living in our care homes and have purchased privately their riser/recliner chair, and this device is not fitted with a lockable handset, we will work with that resident and/or their family to seek a solution and ensure the riser/recliner is replaced by a chair that has a lockable handset. We anticipate this may take the first quarter of 2024 to fully resolve, where the equipment supplied is privately owned and already in use.”

    Source location

    Response from Aria Care
    Page 1 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent action was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Explore and develop a potentially universal lockable cover to restrict access to riser/recliner handset controls.

    Verbatim wording from the response

    “In the meantime, our property services department are working with NH Maintenance (NHM), to explore other options, including the development of a cover that could potentially be used universally, to restrict access to the handset controls, this device would be lockable.”

    Source location

    Response from Aria Care
    Page 1 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent action was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Use compatible Shackleton’s controls to reduce unauthorised access while universal covers are developed.

    Verbatim wording from the response

    “Whilst waiting on the development by NHM, we will look to use the controls provided by Shackleton’s to reduce unauthorised access to riser/recliner chair controls, where they are compatible. For chairs where we cannot control access via the introduction of the covers, we have introduced generic and individualised risk assessments which include the requirement to:”

    Source location

    Response from Aria Care
    Page 2 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent action was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Use generic and resident-specific risk assessments requiring safe chair positioning and electrical isolation where residents cannot independently control chairs.

    Verbatim wording from the response

    “Whilst waiting on the development by NHM, we will look to use the controls provided by Shackleton’s to reduce unauthorised access to riser/recliner chair controls, where they are compatible. For chairs where we cannot control access via the introduction of the covers, we have introduced generic and individualised risk assessments which include the requirement to:”

    Source location

    Response from Aria Care
    Page 2 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent action was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Reduce riser/recliner chair use through a phased approach and purchase assessed-need furniture solely from Shackleton’s with lockable handsets.

    Verbatim wording from the response

    “5) Agreement that via a phased approach Aria Care will reduce the use of riser/recliner chairs across the organisation and where an assessed need exists, riser/recliner furniture will be purchased solely from Shackleton’s, as our furniture partner, with lockable handsets.”

    Source location

    Response from Aria Care
    Page 2 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent action was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Notify the Office for Product Safety and Standards about the incident for consumer-product safety consideration.

    Verbatim wording from the response

    “After receiving your report, HSE contacted the other regulatory bodies who may have an interest or potential regulatory role in respect of this incident. We have written to CQC (who are already in receipt of your report) on the patient safety aspects. HSE has also notified the Office for Product Safety and Standards (OPSS) who are the national regulator for consumer products and who work with the local authority Trading Standards services that undertake most enforcement activity.”

    Source location

    Response from Health and Safety Executives
    Page 2 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent position was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    A universal lockable handset cannot be supplied because riser/recliner chairs use different manufacturers, and some chairs are incompatible with available handsets.

    Verbatim wording from the response

    “Following receipt of the request, Aria Care reached out to Shackleton’s, our furniture supply partner seeking advice and guidance on riser/recliner chairs and the potential for purchase and supply of handsets that could be locked in a secure position and avoid unauthorised access and repositioning.”

    Source location

    Response from Aria Care
    Page 1 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent position was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    A universal lockable handset cover cannot provide an immediate short-term solution because development and trial may take two to three months.

    Verbatim wording from the response

    “In the meantime, our property services department are working with NH Maintenance (NHM), to explore other options, including the development of a cover that could potentially be used universally, to restrict access to the handset controls, this device would be lockable.”

    Source location

    Response from Aria Care
    Page 1 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent position was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Consumer-product safety enforcement for chairs purchasable for domestic use normally falls to local-authority Trading Standards, not HSE.

    Verbatim wording from the response

    “With regard to product safety matters, the chair involved in this tragic incident would appear to meet the definition of a ‘machine’, and therefore the relevant supply legislation would be the Supply of Machinery (Safety) Regulations 2008, which implemented the EU Machinery Directive 2006/42/EC in GB.”

    Source location

    Response from Health and Safety Executives
    Page 2 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent position was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    CQC cannot take further enforcement action because there is insufficient evidence linking the incident directly to a registered person.

    Verbatim wording from the response

    “investigator from CQC’s national enforcement team; investigations are now complete, and it has been adjudged that CQC has no basis to take further action as there is insufficient evidence linking this directly to a registered person. The family were made aware of the outcome of CQC’s decision and CQC will continue to monitor the provider.”

    Source location

    Response from Department of Health and Social Care
    Page 2 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent position was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Service-user safety enforcement in registered care homes is outside HSE’s remit.

    Verbatim wording from the response

    “In England, the Care Quality Commission (CQC) is the lead inspection and enforcement body for safety and quality of treatment and care matters involving patients and service users in receipt of a health or adult social care service from a provider registered with CQC. The Memorandum of Understanding between CQC and HSE explains the respective roles and responsibilities of each organisation with regard to health and safety incidents in the health and adult social care sectors.”

    Source location

    Response from Health and Safety Executives
    Page 1 · response
    Published 6 November 2023

    Open published response

    Source evidence

    How this respondent position was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    CQC is responsible for regulating safety, treatment and care matters involving service users in registered care homes.

    Verbatim wording from the response

    “In England, the Care Quality Commission (CQC) is the lead inspection and enforcement body for safety and quality of treatment and care matters involving patients and service users in receipt of a health or adult social care service from a provider registered with CQC. The Memorandum of Understanding between CQC and HSE explains the respective roles and responsibilities of each organisation with regard to health and safety incidents in the health and adult social care sectors.”

    Source location

    Response from Health and Safety Executives
    Page 1 · response
    Published 6 November 2023

    Open published response
  2. North Yorkshire (West)

    AI-generated summary

    Joseph Francis De Pellegrino-Farrugia · Prevention of Future Deaths report

    This summary was generated using AI from the published report. Please read the original report for the complete account.

    Report summary

    Joseph Francis De Pellegrino-Farrugia, a 92-year-old man with dementia, was injured when his foot became trapped in the mechanism of his electric reclining chair during a transfer to a commode. He sustained crush injuries and ongoing ulceration and necrosis of his left great toe, and died at home on 20 December 2016; the injury contributed to his death. Evidence raised concern that the chair lacked sensors to detect and release a trapped foot.

    Read the report on judiciary.uk

    Source evidence

    How this individual concern was interpreted

    PFD Monitor created a concise, searchable interpretation from the report wording shown below.

    PFD Monitor interpretation

    Lack of chair sensors to detect a foot in the mechanism

    Wider context from the report

    “Evidence was heard to the effect that the chair was not fitted with sensors which would have detected the presence of the foot in the mechanism to prevent the crushing injury and to release the foot. ”

    Source location

    Joseph Francis De Pellegrino-Farrugia · Prevention of Future Deaths report
    Page 2 · concerns

    Open source report

    Source evidence

    How this respondent action was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Fit sensor strips or protective screens beneath chair sides and backs where entrapment risk is identified.

    Verbatim wording from the response

    “AJ Ways can and do fit a sensor strips or a protective screen to the underside of the side and back of the chair to avoid accidental trapping. These are not a mandatory attachment and therefore are only added to the chair in response to orders placed by the retailer who sells the chair to a consumer who identifies that there may be a risk of entrapment to small children and animals.”

    Source location

    2017-0430-Response
    Page 1 · response
    Published 3 July 2017

    Open published response

    Source evidence

    How this respondent position was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Sensors are unsuitable for front legrest entrapment because the legrest flexes, so they cannot address the accident mechanism.

    Verbatim wording from the response

    “Having looked at the evidence presented on how this accident happened I can see that the entrapment occurred at the front of the chair underneath the legrest, the sensors that are used are not suitable for this part of the chair due to the fact that the legrest flexes hence highlighting this in the user instructions.”

    Source location

    2017-0430-Response
    Page 1 · response
    Published 3 July 2017

    Open published response

    Source evidence

    How this respondent position was interpreted

    PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

    PFD Monitor interpretation

    Protective sensors are added only when retailers order them after identifying an entrapment risk.

    Verbatim wording from the response

    “AJ Ways can and do fit a sensor strips or a protective screen to the underside of the side and back of the chair to avoid accidental trapping. These are not a mandatory attachment and therefore are only added to the chair in response to orders placed by the retailer who sells the chair to a consumer who identifies that there may be a risk of entrapment to small children and animals.”

    Source location

    2017-0430-Response
    Page 1 · response
    Published 3 July 2017

    Open published response