Investigation and inquest
On 11 September 2020 I commenced an investigation into the death of Jean DYE aged 77. The investigation concluded at the end of the inquest on 02 July 2025. The conclusion of the inquest was that:
Jean Dye died on 7 September 2020 at Scunthorpe General Hospital as a consequence of an iatrogenic artery dissection which occurred during Percutaneous Coronary Intervention to treat underlying cardiac disease. That was a rare but recognised complication of the procedure. Once the dissection was identified there was a limited opportunity to provide the necessary remedial treatment by the deployment of cardiac stents. Coincidentally at that moment there was a sudden and unexpected failure of electrical power within the treatment room which persisted for approximately 10 minutes. Save that it resulted from an activation of the Emergency Power Off (EPO) circuit, which overrode the back up electrical supply, no clear cause of that loss of power was identified. There was no manual activation of the EPO buttons and a fault within the circuitry was suspected but not established to the required standard. The loss of electrical power removed the ability to provide x ray images and consequently prevented commencement of the stenting procedure until power was regained. Although the stenting was then completed Mrs Dye failed to recover and her death was confirmed. On balance of probabilities, Mrs Dye would have survived but for the loss of electrical power.
Circumstances of the death
Jean Dye died on 7 September 2020 at Scunthorpe General Hospital as a consequence of an iatrogenic artery dissection which occurred during Percutaneous Coronary Intervention to treat underlying cardiac disease. That was a rare but recognised complication of the procedure. Once the dissection was identified there was a limited opportunity to provide the necessary remedial treatment by the deployment of cardiac stents. Coincidentally at that moment there was a sudden and unexpected failure of electrical power within the treatment room which persisted for approximately 10 minutes. Save that it resulted from an activation of the Emergency Power Off (EPO) circuit, which overrode the back up electrical supply, no clear cause of that loss of power was identified. There was no manual activation of the EPO buttons and a fault within the circuitry was suspected but not established to the required standard. The loss of electrical power removed the ability to provide x ray images and consequently prevented commencement of the stenting procedure until power was regained. Although the stenting was then completed Mrs Dye failed to recover and her death was confirmed. On balance of probabilities, Mrs Dye would have survived but for the loss of electrical power.
Coroner’s concerns
The circumstances of Mrs Dye's death are set out above. Her death arose as a consequence of the combination of an iatrogenic injury sustained during a clinical procedure and an untimely loss of power to the Catheter Lab where she was being treated for a period of approximately ten minutes at exactly the time at which her treating consultant required the benefits of real time x ray to facilitate a necessary emergency stenting procedure. The delay whilst power was restored was a critical factor in this death.
The loss of power arose as a result of the Emergency Power Off (EPO) circuit activating. It overrode the emergency power back up system. The reason for that activation was unclear although a physical activation of any of the three EPO buttons was excluded on the evidence. All staff at the scene were unaware of the cause of the loss of power, never having experienced such a situation previously, and an engineer was summoned to attend to reinstate the power, which he did.
There was no light or other indicator within the lab to confirm to those present that the EPO circuit had activated. Likewise there was no restart button within the lab to permit the EPO circuit to be reset. That lay within the plant room elsewhere within the hospital.
I received evidence of a small number of accidental activations of EPO circuits nationally over a 12 month period. It is plainly something which can, and does, occur.
Had staff been aware of the exact cause of the loss of power on this occasion and had they had the opportunity to reset the circuit without the need to await the arrival of an engineer, who inturn had to attend a separate plant room, the downtime would likely have been significantly reduced. Whilst it was not possible to say that the additional time spent on this occasion made a difference between the patient surviving or not, there may well be future cases within which such fine margins are time critical.
I received evidence that there is no current guidance in relation to the siting of such controls remote from the affected room. I invite review of that guidance and of the need for any consequential training.