Issued 28 Sep 2017•Manchester West

Report record

Published report and response evidence

This page connects the concerns raised in this report with statements found in recipients’ published responses. A link shows a clear evidence connection; it does not assign responsibility.

View original report
Concerns
3

Raised in this report

Recipients
4

Named on the report

Responses found
1

Of 4 recipients

Stated actions
0

Described in responses

Source document

Full report text

This is the full text from the original published report.

Open published report

Concerns and recipient responses

Select any concern, action or position to view the source wording.

Report evidence summary

Concerns raised3

  1. Failure to provide nursing staff with instructions on the operational integrity of essential falls-risk equipment
    Part of recurring concern: Failure to ensure clear and followed instructions for safe equipment usePart of recurring concern: Inadequate control of falls risks
  2. Unsuitability of wireless WiFi activation when several fall mats are in proximity
    Part of recurring concern: Unreliable fall-mat safety controls
  3. Unreliability of Rambledguard fall mats for their intended purpose
    Part of recurring concern: Unreliable fall-mat safety controls
Responses linked to these concerns

Each statement is shown once, even when linked to more than one concern.

Respondent positions A position is what a recipient says about a concern when it does not describe a specific action.1

  1. Position

    Wi-Fi is not involved in the fall mats’ processes and therefore cannot cause delays or cross-talk with other devices.

    Stated by Department of Health and Social CareDisputes the concernThe respondent disagreed with part of the concern or the basis for it.

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Failure to provide nursing staff with instructions on the operational integrity of essential falls-risk equipment

Wider context from the report

“The evidence raises the following concerns: 1. The reliability of the Rambledguard fall mats and its fitness for purpose. 2. The suitability of a wireless “WiFi” activated where several fall mats are in place in proximity to each other. 3. Insertion of instructions to nursing staff where the operational integrity of an essential equipment to alleviate falls risks has been identified. ”

Is this part of a recurring concern?

Yes — Failure to ensure clear and followed instructions for safe equipment use; Inadequate control of falls risks.

Open source report

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Unsuitability of wireless WiFi activation when several fall mats are in proximity

Wider context from the report

“The evidence raises the following concerns: 1. The reliability of the Rambledguard fall mats and its fitness for purpose. 2. The suitability of a wireless “WiFi” activated where several fall mats are in place in proximity to each other. 3. Insertion of instructions to nursing staff where the operational integrity of an essential equipment to alleviate falls risks has been identified. ”

Is this part of a recurring concern?

Yes — Unreliable fall-mat safety controls.

Open source report

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Unreliability of Rambledguard fall mats for their intended purpose

Wider context from the report

“The evidence raises the following concerns: 1. The reliability of the Rambledguard fall mats and its fitness for purpose. 2. The suitability of a wireless “WiFi” activated where several fall mats are in place in proximity to each other. 3. Insertion of instructions to nursing staff where the operational integrity of an essential equipment to alleviate falls risks has been identified. ”

Is this part of a recurring concern?

Yes — Unreliable fall-mat safety controls.

Open source report

Source evidence

How this respondent position was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Wi-Fi is not involved in the fall mats’ processes and therefore cannot cause delays or cross-talk with other devices.

Verbatim wording from the response

“Firstly, I understand from Bolton NHS Foundation Trust that it has worked closely with Ramblegard Ltd to explore your concerns. I am advised that there is no issue in regard to Wi-Fi potentially causing a delay, or ‘cross-talking’, with other devices, as Wi-Fi is not involved in the processes.”

Source location

2017-0278-Response-by-Department-of-Health
Page 1 · response
Published 27 November 2017

Open published response

Other statements in published responses

These actions and other statements could not be clearly connected to one concern in this report.

Recipient positions A position is what a recipient says about a concern when they do not describe a specific action.3

  1. 1

    Further regulatory concerns should be directed to Ofcom or Trading Standards, whose remit the product may fall within.

    Stated by Department of Health and Social CareRedirects responsibilityThe respondent said that another organisation was responsible for deciding or taking action.
  2. 2

    The product falls outside the regulatory remit of the Medicines and Healthcare products Regulatory Agency and the Health and Safety Executive.

    Stated by Department of Health and Social CareOutside remitThe respondent said that this matter was outside its role or authority.
  3. 3

    The Trust is responsible for ensuring purchased fall-prevention products comply with necessary delivery requirements.

    Stated by Department of Health and Social CareRedirects responsibilityThe respondent said that another organisation was responsible for deciding or taking action.

Source evidence

How this respondent position was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

Further regulatory concerns should be directed to Ofcom or Trading Standards, whose remit the product may fall within.

Verbatim wording from the response

“My officials have noted that the product information sheet for the Ramblegard Ltd wireless monitor states that it is CE approved and compliant with the Radio and Telecommunications Terminal Equipment (R&TTE) Directive (RED), June 2017. The product may fall within the regulatory remit of Ofcom or Trading Standards and I would suggest contacting them if you have further concerns regarding the product.”

Source location

2017-0278-Response-by-Department-of-Health
Page 2 · response
Published 27 November 2017

Open published response

Source evidence

How this respondent position was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

The product falls outside the regulatory remit of the Medicines and Healthcare products Regulatory Agency and the Health and Safety Executive.

Verbatim wording from the response

“My officials have ascertained that this product does not fall within the regulatory remit of the Medicines and Healthcare products Regulatory Agency or the Health and Safety Executive.”

Source location

2017-0278-Response-by-Department-of-Health
Page 2 · response
Published 27 November 2017

Open published response

Source evidence

How this respondent position was interpreted

PFD Monitor created a concise, searchable interpretation from the published response wording shown below.

PFD Monitor interpretation

The Trust is responsible for ensuring purchased fall-prevention products comply with necessary delivery requirements.

Verbatim wording from the response

“Finally, in liaison with the NHS Shared Business Services (SBS), Departmental officials have confirmed that Ramblegard Ltd products are sold under one of its Framework agreements that enables NHS organisations to contract with suppliers of fall prevention products. Ramblegard Ltd products are supplied by one of the six awarded framework suppliers who have passed the quality criteria in the tender and can be appointed under the Framework agreement. Such quality is also warranted by the supplier to SBS under the terms of the Framework agreement. The Framework gives NHS organisations the ability to purchase goods from qualifying suppliers, pursuant”

Source location

2017-0278-Response-by-Department-of-Health
Page 2 · response
Published 27 November 2017

Open published response
Back to top