This is the full text from the original published report.

Open published report

Concerns and recipient responses

Select any concern, action or position to view the source wording.

Report evidence summary

Concerns raised5

  1. Remounting of the valve contrary to the manufacturer’s instructions for use
    Part of recurring concern: Failure to ensure clear and followed instructions for safe equipment use
  2. Failure to discuss and record the rationale for departing from the manufacturer’s instructions for use
    Part of recurring concern: Failure to reliably document the rationale for consequential decisions
  3. Lack of awareness among the primary surgeon and operating team of the risks of departing from the manufacturer’s instructions for use
    Part of recurring concern: Failure to ensure clear and followed instructions for safe equipment use
Responses linked to these concerns

Each statement is shown once, even when linked to more than one concern.

No linked response statements

No respondent-stated action or position is clearly linked to these concerns.

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Remounting of the valve contrary to the manufacturer’s instructions for use

Wider context from the report

“(1) During Mrs Hynes operation a direction was given to remount the Sorin Carbomedics 23mm mechanical aortic valve on its holder whilst preparations to implant the valve were undertaken. Remounting the valve on its holder is contrary to the manufacturer’s instructions for use. Concerns arising are: a) The rational for departing from the manufacturer's instructions for use was neither discussed nor recorded. b) The primary surgeon and operating team were unaware of the risks of departing from the manufacturer's instructions for use namely potential inverted remount. c) A scrub nurse with neither training nor experience was instructed to remount the valve contrary to the manufacturer's instructions for use. The rational for this direction was neither discussed nor recorded. d) The primary surgeon with overall responsibility for the procedure did not instruct remounting of the valve; There are no recorded discussions with the primary surgeon on this issue. ”

Is this part of a recurring concern?

Yes — Failure to ensure clear and followed instructions for safe equipment use.

Open source report

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Failure to discuss and record the rationale for departing from the manufacturer’s instructions for use

Wider context from the report

“(1) During Mrs Hynes operation a direction was given to remount the Sorin Carbomedics 23mm mechanical aortic valve on its holder whilst preparations to implant the valve were undertaken. Remounting the valve on its holder is contrary to the manufacturer’s instructions for use. Concerns arising are: a) The rational for departing from the manufacturer's instructions for use was neither discussed nor recorded. b) The primary surgeon and operating team were unaware of the risks of departing from the manufacturer's instructions for use namely potential inverted remount. c) A scrub nurse with neither training nor experience was instructed to remount the valve contrary to the manufacturer's instructions for use. The rational for this direction was neither discussed nor recorded. d) The primary surgeon with overall responsibility for the procedure did not instruct remounting of the valve; There are no recorded discussions with the primary surgeon on this issue. ”

Is this part of a recurring concern?

Yes — Failure to reliably document the rationale for consequential decisions.

Open source report

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Lack of awareness among the primary surgeon and operating team of the risks of departing from the manufacturer’s instructions for use

Wider context from the report

“(1) During Mrs Hynes operation a direction was given to remount the Sorin Carbomedics 23mm mechanical aortic valve on its holder whilst preparations to implant the valve were undertaken. Remounting the valve on its holder is contrary to the manufacturer’s instructions for use. Concerns arising are: a) The rational for departing from the manufacturer's instructions for use was neither discussed nor recorded. b) The primary surgeon and operating team were unaware of the risks of departing from the manufacturer's instructions for use namely potential inverted remount. c) A scrub nurse with neither training nor experience was instructed to remount the valve contrary to the manufacturer's instructions for use. The rational for this direction was neither discussed nor recorded. d) The primary surgeon with overall responsibility for the procedure did not instruct remounting of the valve; There are no recorded discussions with the primary surgeon on this issue. ”

Is this part of a recurring concern?

Yes — Failure to ensure clear and followed instructions for safe equipment use.

Open source report

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Failure to involve and record discussions with the primary surgeon before remounting the valve

Wider context from the report

“(1) During Mrs Hynes operation a direction was given to remount the Sorin Carbomedics 23mm mechanical aortic valve on its holder whilst preparations to implant the valve were undertaken. Remounting the valve on its holder is contrary to the manufacturer’s instructions for use. Concerns arising are: a) The rational for departing from the manufacturer's instructions for use was neither discussed nor recorded. b) The primary surgeon and operating team were unaware of the risks of departing from the manufacturer's instructions for use namely potential inverted remount. c) A scrub nurse with neither training nor experience was instructed to remount the valve contrary to the manufacturer's instructions for use. The rational for this direction was neither discussed nor recorded. d) The primary surgeon with overall responsibility for the procedure did not instruct remounting of the valve; There are no recorded discussions with the primary surgeon on this issue. ”

Is this part of a recurring concern?

Yes — Failure to provide effective senior clinical oversight of patient care; Incomplete, inaccurate or unavailable clinical and care records; Unreliable communication of patient-care information between clinical staff; Unreliable documentation and communication of shared clinical decisions.

Open source report

Source evidence

How this individual concern was interpreted

PFD Monitor created a concise, searchable interpretation from the report wording shown below. Response links show a clear evidence connection; they do not assign responsibility.

PFD Monitor interpretation

Assignment of valve remounting to an untrained and inexperienced scrub nurse

Wider context from the report

“(1) During Mrs Hynes operation a direction was given to remount the Sorin Carbomedics 23mm mechanical aortic valve on its holder whilst preparations to implant the valve were undertaken. Remounting the valve on its holder is contrary to the manufacturer’s instructions for use. Concerns arising are: a) The rational for departing from the manufacturer's instructions for use was neither discussed nor recorded. b) The primary surgeon and operating team were unaware of the risks of departing from the manufacturer's instructions for use namely potential inverted remount. c) A scrub nurse with neither training nor experience was instructed to remount the valve contrary to the manufacturer's instructions for use. The rational for this direction was neither discussed nor recorded. d) The primary surgeon with overall responsibility for the procedure did not instruct remounting of the valve; There are no recorded discussions with the primary surgeon on this issue. ”

Is this part of a recurring concern?

Yes — Unsafe assignment of staff without the required qualifications or competence to care work.

Open source report